11 | Keysight | InniiVision 6000L Series Low-Prole Oscilloscopes - Data Sheet
Performance characteristics (continued)
FFT
Points Fixed at 1000 points
Source of FFT Scope channels 1, 2, 3 or 4, 1+2, 1-2, 1x2
Window Rectangular, flattop, Hanning
Noise floor –50 to –90 dB depending on averaging
Amplitude Display in dBV, dBm at 50 Ω
Frequency resolution 0.05/(time per div)
Maximum frequency 50/(time per div)
Storage
Save/recall (non-volatile) 10 setups and traces can be saved and recalled internally. Secure environment mode (-SEC)
ensures setups and traces are stored to volatile memory.
Storage type and format USB 1.1 drive on front (/drive0) and rear (/drive5) panels
Image formats: BMP (8 bit), BMP (24 bit) and PNG (24 bit)
Data formats: X and Y (time/voltage) values in CSV, ASCII XY and binary format
Trace/setup formats: Recalled
I/O
Standard ports USB 2.0 high speed, 10/100-BaseT LAN, IEEE488.2 GPIB, XGA video output
Max transfer rate IEEE488.2 GPIB: 500 kbytes/sec
USB (USBTMC-USB488): 3.5 Mbytes/sec
100 Mbps LAN (TCP/IP): 1 Mbytes/sec
Remote front panel
Built-in help language support for English, German, French, Russian, Japanese, Traditional
Chinese, Simplified Chines, Korean, Spanish, Portuguese and Italian
Throughput of scope channels 100,000 waveforms/sec in real-time mode to remote monitor
Resolution of video output XGA
Waveform controls Waveform intensity of 256 levels, vectors on/off, infinite persistence on/off
General characteristics
Rack mounting Supplied with all necessary hardware (except tools) for installation into a standard EIA 19-inch
rack
Physical size 43.5 cm W x 27 cm D x 4.2 cm H (without brackets)
Weight Net: 2.45 kg (5.4 lbs.)
Shipping: 6.2 kg (13.6 lbs.)
Probe comp output Frequency ~1.2 kHz
Amplitude ~2.5 V